Predischarge screening of neonates with transcutaneous bilirubinometer: Comparison of various nomograms
Thrilok Natarajan1, Ramesh Srinivasan2, Sumathi Raju2, Suvetha Kannapan3
1 Department of Pediatrics, Queen Alexandra Hospital, Portsmouth, England, UK 2 Department of Pediatrics, PSG Institute of Medical Sciences and Research, Coimbatore, Tamil Nadu, India 3 Department of Community Medicine, PSG Institute of Medical Sciences and Research, Coimbatore, Tamil Nadu, India
Correspondence Address:
Ramesh Srinivasan Department of Pediatrics, PSG Institute of Medical Sciences and Research, Coimbatore, Tamil Nadu India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/jcn.jcn_129_21
|
|
Objective: To compare the diagnostic accuracy of various transcutaneous bilirubin (TcB) nomograms for predischarge screening. Methods: The paired total serum bilirubin (TSB) and TcB measurements collected in neonates ≥35 weeks and ≥2000 g birth weight were analyzed. BiliCare™ bilirubinometer was used for TcB measurement. We chose the following nomograms for the study: Bhutani nomogram, Maisel's nomogram, Agarwal nomogram, Thakkar nomogram, American Academy of Pediatrics (AAP) nomogram within 3 mg/dl of phototherapy cutoff, AAP nomogram >70% of phototherapy cutoff and if TcB value is above 13 mg/dl. The diagnostic accuracy of these nomograms for TcB was compared with TSB plotted in the Bhutani nomogram. Results: TcB showed a positive correlation with TSB (Pearson correlation coefficient = 0.783). Bhutani nomogram, Maisel's nomogram and AAP (using within 3 mg/dL cutoff) nomogram showed good sensitivity and low false-negative rate while avoiding blood draws in most neonates. Conclusion: Bhutani nomogram, Maisel's nomogram, and AAP (using within 3 mg/dL of phototherapy cutoff) nomograms have comparable diagnostic accuracy for predischarge bilirubin screening in neonates.
|
|
|
|
[FULL TEXT] [PDF]* |
|
 |
|